21 February 2022

9 Important Things to Know When Making Sensitive Measurements with Oscilloscopes

We've routinely posted on how you can characterize your total measurement system to gain important "situational awareness" when using an oscilloscope to make sensitive measurements. The knowledge gained from these tests helps you properly interpret your measurement results so that you can deduce what is actually going on with your circuit, versus what is an artifact of the measurement system. Listed here are nine important things you should know before making sensitive measurements with your oscilloscope, with links to blog posts that instruct you how to test them.

19 July 2021

How to Test Noisy Power Supply Outputs

Figure 1: 3.3 V output of a DC-DC converter. The waveform shows the nominal DC level, ripple and high frequency noise bursts.
Figure 1: 3.3 V output of a DC-DC converter.
The waveform shows the nominal DC level,
ripple and high frequency noise bursts.
Did you ever acquire the output of a power supply with your oscilloscope and find an unexpectedly high level of noise? Did you try adding filter capacitors only to find the noise level was not changed? 

In this post, we'll discuss how the choice of probe affects the noise present in power measurements, as well as how oscilloscope settings such as termination impedance, bandwidth and coupling can be adjusted to lessen noise and improve measurement results.

Figure 1 shows a typical DC-DC converter output measurement. The mean value of the waveform is 3.294 V.  Ripple appears at the switching frequency of 1.2 MHz, and noise in the form of high frequency bursts and baseline thickening is visible throughout.

Waveforms like this can be acquired with a 10:1 high impedance probe, a 1:1 coaxial cable connection, or a 1.2:1 rail probe using either DC or AC coupling, as available.  Figure 2 summarizes how each oscilloscope/probe configuration affects the measurement.

18 January 2021

Situational Awareness: Testing Oscilloscope Outer Limits

Fig 1. 40 ps signal measured full bandwidth on a 1 GHz oscilloscope shows visible over/undershoot.
Fig 1. 40 ps signal measured full bandwidth on a
1 GHz oscilloscope shows visible over/undershoot.
Nothing is perfect. Every test instrument has its limits, and knowing the limits to your oscilloscope’s bandwidth in response to real-world signals helps to develop situational awareness when making measurements. This is especially true when testing signals that are at or very near the specified bandwidth limit of the instrument.

The measurements we’ll demonstrate were made on a WaveSurfer 4104HD, a 12-bit, 4-channel, 1 GHz bandwidth oscilloscope that samples at up to 5 GS/s.

06 December 2017

Probing Techniques and Tradeoffs (Part IV)

Applying bandwidth filters to a 2.5-GHz clock signal clearly shows the effect of bandwidth on rise time
Figure 1: Applying bandwidth filters to a 2.5-GHz clock
signal clearly shows the effect of bandwidth on rise time
The topic of probe bandwidth is a broad and deep one. We began our discussion of bandwidth in an earlier post with some basic information about what bandwidth means and the importance of the -3 dB point. Next, we looked at a Fourier deconstruction of a square wave into its fundamental and the lower-order harmonics, and covered the importance of bandwidth in capturing enough harmonic content to understand the signal's overall shape.