You need to test, we're here to help.

You need to test, we're here to help.
Showing posts with label situational awareness. Show all posts
Showing posts with label situational awareness. Show all posts

22 February 2021

Don't Attach Multiple Probes to the Same Place at the Same Time!

Figure 1. Response of two different probes to an upper-side gate drive measurement.
Figure 1. Response of two different probes to an
upper-side gate drive measurement.
Along with using single-ended passive probes for high-voltage measurements, another probing "no no" to avoid is attaching multiple probes to the same place at the same time.

You are probably aware that all measuring instruments, including oscilloscopes, are subject to conditions of observability.  As we discussed in a recent post on The Impact of the Interconnect, the very act of connecting the oscilloscope to the circuit with a particular probe affects the measurement in a particular way. Probes affect the circuit by applying additional resistance and capacitance loads in parallel with the circuit at the test point. Moreover, the probes themselves limit the fidelity of the measurement due to limits of bandwidth, slew rate and common mode response. So, it is always a good idea when making a measurement to compare how different probes/interconnects will affect the measurement…but don’t try to do it all at once.  

01 February 2021

Situational Awareness: RF Noise in the Lab

Fig. 1. Time domain (top) and spectral (bottom) views of signal shown in SPECTRUM-1 on a WaveSurfer 4000HD.
Fig. 1. Time domain (top) and spectral (bottom) views of
s
ignal shown in SPECTRUM-1 on a WaveSurfer 4000HD. 
Laboratories have multiple sources of RF that can affect measurements, like computers, cell phones, routers, local radio and TV stations, even a nearby airport. Knowing the RF background of your lab is another part of Situational Awareness. Only by knowing the background can you know what is actually due to the device under test.

One way to read RF is through spectral analysis of Fourier transforms (DFT and FFT). FFTs take a time domain view of a signal (e.g, amplitude versus time trace) and change it into a spectrum of amplitude plotted as a function of frequency. Frequency spectrums are great for observing signals than are asynchronous with the process being measured. They have a lower noise floor and offer better dynamic range than do time domain plots.  Consider the views of the same signal shown in the time domain and frequency domain in Fig. 1.

18 January 2021

Situational Awareness: Testing Oscilloscope Outer Limits

Fig 1. 40 ps signal measured full bandwidth on a 1 GHz oscilloscope shows visible over/undershoot.
Fig 1. 40 ps signal measured full bandwidth on a
1 GHz oscilloscope shows visible over/undershoot.
Nothing is perfect. Every test instrument has its limits, and knowing the limits to your oscilloscope’s bandwidth in response to real-world signals helps to develop situational awareness when making measurements. This is especially true when testing signals that are at or very near the specified bandwidth limit of the instrument.

The measurements we’ll demonstrate were made on a WaveSurfer 4104HD, a 12-bit, 4-channel, 1 GHz bandwidth oscilloscope that samples at up to 5 GS/s.

11 January 2021

Four Measurement Best Practices

Questions for taking measurements
To start the New Year right, we’re going to talk about four measurement "best practices", which will help you get the most out of any oscilloscope you have. These are important when doing any type of measurement—and you can get a good start on them simply by asking yourself the four questions in the sidebar.

1. Anticipate the results

Those who are familiar with Dr. Eric Bogatin’s Rule #9 will know this one. Before you do any measurement, anticipate what you expect the result to be, because that is the most important way of identifying if there is a potential problem.