| Figure 1: The typical EFT test signal consists of multiple exponential pulses arranged as pulse bursts. |
![]() |
| Figure 2: Acquisition of two EFT bursts at 1.25 GS/s, zoomed to show several timing epochs. |
![]() |
| Figure 1: The typical EFT test signal consists of multiple exponential pulses arranged as pulse bursts. |
![]() |
| Figure 2: Acquisition of two EFT bursts at 1.25 GS/s, zoomed to show several timing epochs. |
![]() |
| Figure 1: A typical EMC surge test waveform with 1.2 µs rise time and 50 µs half amplitude time. Zoom trace Z1 shows the details of the rise time. Click any image to expand it. |
Surge pulses are similar to ESD pulses in that they have a very fast rise time, but the fall time is much, much slower. Figure 1 shows a typical surge pulse waveform. Surge testing involves similar measurements to those made for ESD pulse testing—such as rise time, pulse width and max—but surge testing also requires some additional measurements, such as area under the pulse curve and transmitted charge.
![]() |
| Figure 2: To verify the surge generator waveform, the oscilloscope is connected to the generator through an attenuator. |
Following are three, important things to do to make sure you get the best surge measurements from your oscilloscope.
![]() |
| Figure 1: An ESD calibration test setup. The ESD gun discharges its waveform into a properly attenuated current target. |
Oscilloscopes are most often used to “test the tester” in ESD pulse test setups, confirming that the pulse from the ESD gun is the right shape and meets the requirements of the standard to which the device is being tested. A typical calibration test setup is shown in Figure 1. The pulse from the ESD gun is fired directly into a current shunt target connected to the oscilloscope through an attenuator required to keep the signal within the limits of the oscilloscope’s 50 Ω input, which is used for this testing. Then, key parameters of the ESD pulse are measured per one of several standards, such as IEC 61000-4-2.
ESD standards require a range of measurements. The most common are the initial edge 10% to 90% rise time, peak amplitude, pulse width, amplitude and current levels at specified times from the initial edge (e.g., T1 and T2), and time to half value.
Following are four, important things to do to make sure you get the best ESD pulse measurements from your oscilloscope.
![]() |
| Figure 1: Parameter math functions provide a way to create custom parameters |
![]() |
| Figure 1: The track math function shows how data changes over time |
![]() |
| Figure 1: Cursors (top) and parameter measurements (bottom) are both powerful tools in their own right |
![]() |
| Figure 1: Measure tools are at the heart of an oscilloscope's utility |
Figure 1: Skew between voltage and current probes
results in power measurement errors
|