| Figure 1: A typical EMC surge test waveform with 1.2 µs rise time and 50 µs half amplitude time. Zoom trace Z1 shows the details of the rise time. Click any image to expand it. |
Surge pulses are similar to ESD pulses in that they have a very fast rise time, but the fall time is much, much slower. Figure 1 shows a typical surge pulse waveform. Surge testing involves similar measurements to those made for ESD pulse testing—such as rise time, pulse width and max—but surge testing also requires some additional measurements, such as area under the pulse curve and transmitted charge.
![]() |
| Figure 2: To verify the surge generator waveform, the oscilloscope is connected to the generator through an attenuator. |
Following are three, important things to do to make sure you get the best surge measurements from your oscilloscope.