You need to test, we're here to help.

You need to test, we're here to help.
Showing posts with label WavePro HD. Show all posts
Showing posts with label WavePro HD. Show all posts

08 June 2020

Charting High-Resolution Oscilloscope Performance


Eric Bogatin, Signal Integrity Evangelist, Teledyne LeCroy

Signal measured by 8-bit and 12-bit oscilloscope.
Figure 1. 8-bit vs. 12-bit oscilloscope waveforms.
There are hundreds of different oscilloscope models from a dozen different vendors. How do you make sense of all of their different features to pick the one instrument right for your application?  Here is a simple way of comparing high-resolution oscilloscopes.

High-resolution usually refers to the vertical resolution, the quantization of a waveform into a fixed number of vertical levels measured by the oscilloscope's Analog-to-Digital Converter (ADC). The more bits of vertical resolution, the more levels and the more detailed the waveform rendered. The earliest digital storage oscilloscopes (DSO) started as 8-bit resolution, or 256 vertical levels. But in the last ten years, as ADC chip technology got faster, higher resolution appeared in the industry, pioneered by Teledyne LeCroy.  Now, many oscilloscopes come with 10-bit and 12-bit vertical resolution. Why would you want higher resolution? The answer: the ability to see lower level signals and better dynamic range. An example of the difference between the same signal measured with an 8-bit and 12-bit oscilloscope is shown in Figure 1. 

08 April 2019

Fast Fourier Transforms: Automatic Edition

Figure 1: Shown is the user interface for Teledyne LeCroy's Spectrum Analyzer software option
Figure 1: Shown is the user interface for Teledyne LeCroy's
Spectrum Analyzer software option
Many motorists love the experience of driving a sporty car with its convertible top down and a stickshift manual transmission. Others don't want the hassle of a clutch and prefer the user-friendly feel of a slick automatic that does some of the work for them. Oscilloscopes can provide the same sort of choice for many measurement tasks in the same instrument.

25 March 2019

The Resolution Revolution In Oscilloscopes

Figure 1: Teledyne LeCroy's WavePro HD exemplifies today's high-resolution, high-bandwidth instruments
Figure 1: Teledyne LeCroy's WavePro HD exemplifies
today's high-resolution, high-bandwidth instruments
Oscilloscopes have been around for a very long time now, and older oscilloscope users will remember the heyday of those analog boat anchors of the '60s and '70s. Many have survived and are still usable if you don't need much bandwidth (and have the means to calibrate them if necessary). I used to scout local hamfests looking for bargains on them. Many techs and engineers cut their teeth on those behemoths. You could learn a lot about design if you poked around inside them, too.

08 November 2018

How Tip Inductance Impacts a Probing System's Bandwidth

Shown are FFT plots of a 10-MHz, fast-edge square wave reaching the oscilloscope via direct coax connection  (orange-yellow plot) and 10x passive probe fitted with a coax tip adapter (straw-colored plot)
Figure 1: Shown are FFT plots of a 10-MHz, fast-edge square
wave reaching the oscilloscope via direct coax connection
(orange-yellow plot) and 10x passive probe fitted with a coax
tip adapter (straw-colored plot)
If you're using 10x passive probes with your oscilloscope, it's important to understand the bandwidth of your probing system and how it's affected by various methods of probing the signal of interest. There's a relatively easy way to determine this parameter by probing a fast-edge, 10-MHz signal from a square-wave generator. Doing so can also instruct us in the effects of tip inductance on the probe's bandwidth.

29 August 2018

A Walk-Through of Ground-Bounce Measurements

The trigger pulse from the MCU is one clock cycle in width
Figure 1: The trigger pulse from the
MCU is one clock cycle in width
In earlier posts in this series, we've explained what ground bounce is and how it happens. We have also taken a deeper dive into the use of I/O drivers to implement sense lines that let us better quantify and analyze what kind of ground-bounce hit our system is taking. Now, let's look at a detailed example of how to measure and diagnose ground bounce.