08 November 2018

How Tip Inductance Impacts a Probing System's Bandwidth

Shown are FFT plots of a 10-MHz, fast-edge square wave reaching the oscilloscope via direct coax connection  (orange-yellow plot) and 10x passive probe fitted with a coax tip adapter (straw-colored plot)
Figure 1: Shown are FFT plots of a 10-MHz, fast-edge square
wave reaching the oscilloscope via direct coax connection
(orange-yellow plot) and 10x passive probe fitted with a coax
tip adapter (straw-colored plot)
If you're using 10x passive probes with your oscilloscope, it's important to understand the bandwidth of your probing system and how it's affected by various methods of probing the signal of interest. There's a relatively easy way to determine this parameter by probing a fast-edge, 10-MHz signal from a square-wave generator. Doing so can also instruct us in the effects of tip inductance on the probe's bandwidth.

10 October 2018

Putting Probes in Perspective

Probe, cable, and oscilloscope form a system that makes or breaks the accuracy of signal acquisitions
Figure 1: Probe, cable, and oscilloscope form a system
that makes or breaks the accuracy of signal acquisitions
Few aspects of using an oscilloscope are as important as the probe: after all, the probe forms both the mechanical and electrical interfaces between the device under test (DUT) and the oscilloscope itself. To feed a signal into an oscilloscope, we're limited to a coaxial connection. Thus, we need a geometry transformer that picks up the signal of interest from the DUT and transfers it to the oscilloscope's coaxial connection.

27 November 2017

Probing Techniques and Tradeoffs (Part III)

Bandwidth is defined as the frequency at which the ratio of the displayed amplitude to the input amplitude is -3 dB (or 0.707)
Figure 1: Bandwidth is defined as the frequency at which
the ratio of the displayed amplitude to the input amplitude
is -3 dB (or 0.707)
Any discussion of oscilloscopes and/or probes must include the topic of analog bandwidth. Bandwidth is one of a short list of key specifications for a testbench setup. All oscilloscopes and probes come to market with a bandwidth specification, which is defined as:

The frequency at which the ratio of the displayed amplitude to the input amplitude is -3 dB (or 0.707).

This is known as the "-3 dB point," or the half-power point (Figure 1). At this frequency, a sine-wave input signal is attenuated to 70.7% of its true amplitude. Any higher frequencies will likely be distorted on the display, making accurate measurements and calibration impossible.

01 April 2013

Oscilloscope Basics: Oscilloscope Bandwidth

Among the most important basic specifications of a digital oscilloscope is its bandwidth. Knowing a bit about bandwidth and the influences on the specification can be very helpful in selecting the right oscilloscope for your application. This post will cover some fundamental aspects of oscilloscope bandwidth.