You need to test, we're here to help.

You need to test, we're here to help.
Showing posts with label TDR. Show all posts
Showing posts with label TDR. Show all posts

05 July 2022

A Tale of Two Calibrations: Vector Network Analyzer vs. WavePulser 40iX

Figure 1: This sequence diagram of the classic SOLT 2-path calibration shows the order of connections required.
Figure 1: This sequence diagram of the
classic SOLT 2-path calibration shows
the order of connections required. 
It was the best of S-parameter measurements, it was the worst of S-parameter measurements…and the difference was in the calibration.  Calibrating a vector network analyzer (VNA) before making any measurements is required in order to reduce errors from imperfect channel matching, less than optimal directivity in the directional couplers and cable response issues. While VNAs are precisely calibrated at the factory, that calibration only extends to the front panel measurement ports. There will inevitably be drift on the internal paths over time. Also, any cables, adaptors or fixtures connected to the measurement ports must be characterized and de-embedded in order to make exact measurements of the device under test (DUT).  

There are many possible calibration methods depending on the number of ports and paths being measured.  For simplicity, let’s consider the common 2-port, 2-path calibration.  This calibration method will yield a full set of S-parameters for the two ports: S11, S12, S21 and S22.  It requires the use of a short, open, load and through (SOLT) calibration reference standard, along with the cables used in the test setup, as shown in Figure 1.

14 March 2022

WavePulser 40iX vs. Time Domain Reflectomer (TDR) or Vector Network Analyzer (VNA)

Figure 1: S-parameters and TDR responses are two means to the same end, characterizing interconnects.
Figure 1: S-parameters and TDR responses are two means to
the same end, characterizing interconnects.
Oscilloscopes are used to measure signals, usually as voltages vs. time, and signals come from active devices. But interconnects are passive structures that don’t produce their own signals. To characterize interconnects, you need a stimulus-response system.

There are two, principal types of stimulus-response systems used to characterize interconnects: Vector Network Analyzers (VNAs) used to measure S-parameters in the frequency domain, and Time Domain Reflectometers used to measure impulse responses in the time domain. Each uses a different type of incident signal and a different formalism, but as long as the interconnect is linear, passive and time invariant, both S-parameters and impulse responses yield the same information content in different formats and can be translated from one into another.

So, which do you need? We’ll look briefly at what each does and what are the criteria that might require you to have one versus the other.

21 June 2021

Automotive Ethernet MDI S-parameter Testing

Figure 1: MDI S-parameter tests treat the Base-T1 pair as a balance transmission line and check that reflections don't cause either excessive power loss or mode conversion that can disrupt the signal.
Figure 1: MDI S-parameter tests treat the Base-T1
pair as a balance transmission line and check that
reflections don't cause either excessive power loss
or mode conversion that can disrupt the signal.
As said earlier, the automotive industry has very stringent EMC/EMI requirements, and all Automotive Ethernet standards are designed to ensure good operation even in the presence of high EMI. Not only is there the potential for interference from all the different electronic systems within the vehicle, nothing is stopping you from parking your vehicle below high-voltage transmission wires or in other high EMI fields. 

For this reason, all Automotive Ethernet standards have defined S-parameter tests to be performed at the Medium Dependent Interface (MDI). The assumption is that the single twisted pair that is the basis for all Base-T1 transmissions can be treated as a balanced, differential transmission line with some crosstalk. It is a very real-world application of S-parameters, which can seem so academic.

Two, mixed-mode S-parameters are measured at the MDI reference plane. The tests ensure that there is neither too much loss of power from reflections, nor too much mode conversion into differential signal, that it will disrupt the information of the PAM3 encoded signal.

17 May 2021

Converting Single to Mixed-Mode S-Parameters

Figure 1: Model of two transmission lines with crosstalk showing the transmission and crosstalk related S-parameters.
Figure 1: Model of two transmission lines with crosstalk
showing the transmission and crosstalk related S-parameters.

We have introduced mixed mode S-parameters and developed a formal structure for handling them. It is now time to discuss converting single-ended S-parameters into mixed-mode S-parameters. This is important because every instrument manufacturer obtains mixed mode S-parameters by first measuring single-ended S-parameters, then converting them mathematically to mixed-mode. This assumes that the interconnects being measured are passive, linear and time invariant.  Let’s begin with our model of two transmission lines with crosstalk shown in Figure 1.

24 August 2020

How to Connect the Returns to an Unshielded Twisted Pair

Figure 1. Three options for connecting a TDR to a UTP.
Figure 1. Three options for connecting a TDR to a UTP.
During a recent webinar, Dr. Eric Bogatin was asked several questions about how to measure unshielded twisted pairs (UTPs), which are differential pairs with no return path. Here is his answer to one of the questions. For more information, check out the whole webinar on Differential Pairs with No Return Paths.

Q: For the UTP measurement, do you need to connect the SMA connector shield grounds together?

A: Yes. It’s very important to make sure the grounds are connected at the cable connectors.

There are really three ways the connections can be made from a differential TDR to the UTP cable. These three options are shown clockwise from left to right in Figure 1.

17 December 2014

What S-parameters Reveal About Interconnects (Part III)

How ripple is introduced into S11 and S21
Figure 1: How ripple is introduced into S11 and S21
S-parameters are a great tool for understanding exactly what happens to a signal as it traverses an interconnect such as a transmission line. How much of it propagates through, and how much reflects off of impedance mismatches? From plotting return loss against insertion loss, we've weighed how much return loss may be tolerable before it significantly impacts insertion loss. Now we'll turn our attention to some common patterns exhibited by S11 and S21 and what they mean to the performance of an interconnect.

09 December 2014

What S-parameters Reveal About Interconnects (Part II)

Measuring S-parameters of a two-port interconnect
Figure 1: Measuring S-parameters
of a two-port interconnect
Having previously covered some of the fundamentals of S-parameters, it's now time to dig a little deeper into what they can show us about an interconnect; say, for example, a two-port microstrip line on a PC board. Unlike the one-port DUT in our earlier post, this configuration gives us the opportunity to look at not only S11 (return loss or reflected signal), but also S21 (insertion loss or transmitted signal).

03 December 2014

What S-Parameters Reveal About Interconnects

S-parameters are derived by applying an incident wave to an interconnect
Figure 1: S-parameters are derived by applying an incident
wave to an interconnect; we can consider this process in either
the time or frequency domains
S-parameters are a popular means of characterizing an interconnect. By feeding the interconnect with a precision reference signal and measuring how much of that signal propagates through the connector and how much is reflected, we learn everything we need to know about its performance. This will be the first in a series of posts about the insights we can glean from S-parameters with practical examples of common measurement scenarios.

15 May 2014

Back to Basics: S-parameters

S-matrices for one-, two-, and three-port RF networks
Figure 1: S-matrices for one-, two-,
and three-port RF networks
Suppose you have an optical lens of some sort onto which you shine a light with a known photonic output. While most of the incident light passes through the lens, some fraction of the light is reflected and some is absorbed (the behavior is also dependent on the wavelength of the incident light). You'd like to characterize that lens: Exactly how much light was reflected? How much passed through? What is it about the lens that prevented all of the light from passing through?