05 July 2022

A Tale of Two Calibrations: Vector Network Analyzer vs. WavePulser 40iX

Figure 1: This sequence diagram of the classic SOLT 2-path calibration shows the order of connections required.
Figure 1: This sequence diagram of the
classic SOLT 2-path calibration shows
the order of connections required. 
It was the best of S-parameter measurements, it was the worst of S-parameter measurements…and the difference was in the calibration.  Calibrating a vector network analyzer (VNA) before making any measurements is required in order to reduce errors from imperfect channel matching, less than optimal directivity in the directional couplers and cable response issues. While VNAs are precisely calibrated at the factory, that calibration only extends to the front panel measurement ports. There will inevitably be drift on the internal paths over time. Also, any cables, adaptors or fixtures connected to the measurement ports must be characterized and de-embedded in order to make exact measurements of the device under test (DUT).  

There are many possible calibration methods depending on the number of ports and paths being measured.  For simplicity, let’s consider the common 2-port, 2-path calibration.  This calibration method will yield a full set of S-parameters for the two ports: S11, S12, S21 and S22.  It requires the use of a short, open, load and through (SOLT) calibration reference standard, along with the cables used in the test setup, as shown in Figure 1.

13 June 2022

Oscilloscope Basics: Cal Out and Aux Out

Figure 5. New PCIe 6.0 compliance pattern signal.
Fig. 1: Cal Out and Aux Out 
provide many useful outputs.
Oscilloscopes are generally thought of in terms of the signals that are input to them, but even oscilloscopes that are not equipped with function/signal generators can usually output some useful test signals.

Nearly all oscilloscopes have a Cal Out (calibration output) terminal on the front. Most Teledyne LeCroy oscilloscopes also have an Aux Out (auxiliary output) connector on either the front or back, depending on  model. Both outputs provide configurable signals that can assist you to compensate probes and attenuators, test frequency response, trigger waveform acquisition and coordinate multiple test instruments.

17 August 2017

Back to Basics: Metrology

Some sciences are more obscure than others; these may be of interest and importance to relatively few. The term “metrology” may leave even some science geeks scratching their heads, but for those in the test and measurement arena, metrology is an extremely critical area of scientific endeavor.