04 January 2022

Oscilloscope Basics: Stabilizing Waveform Display, Pt. 1

Figure 1: A free running oscilloscope starts each acquisition at a different point on the waveform, resulting in an unstable display.  A triggered oscilloscope starts each acquisition at the same point on the waveform, resulting in a stable display.
Figure 1: A free running oscilloscope starts each
acquisition at a different point on the waveform,
resulting in an unstable display.  A triggered oscilloscope
starts each acquisition at the same point on the
waveform, resulting in a stable display. 
An unsynchronized, unstable oscilloscope display is useless for making measurements, but proper triggering can synchronize the oscilloscope sample clock to specific waveform events so that the acquired waveforms appear stable on the display.  Let’s look at why signals can appear unstable and what to do about it.  

Oscilloscopes are sampling devices; they sample the incoming signal at a uniform rate.  The timing of a signal applied to the input of an oscilloscope is most probably asynchronous with the oscilloscope’s sampling clock.  If the oscilloscope timebase is allowed to run free—that is, not synchronized to the timing of the input signal—then each oscilloscope acquisition potentially begins at a different point on the input waveform, as shown in Figure 1.

13 May 2016

The Evolving User Interface: Cursors, Triggers, and More

Need to reposition a cursor? Touch, hold, and drag it
Figure 1: Need to reposition a
cursor? Touch, hold, and drag it
In our review of the evolving user interface as embodied in Teledyne LeCroy's MAUI with OneTouch gesture control, we've seen how touch, drag, swipe, pinch, and flick gestures have combined to transform the oscilloscope UI into something that enhances and encourages creativity. Here, we'll cover a few of the remaining features of OneTouch, showing how the concept permeates the MAUI UI in the recently launched WaveRunner 8000 series of instruments.

18 December 2013

Oscilloscope Basics: Trigger Holdoff

As discussed in an earlier post, triggering is the means by which we can coax an oscilloscope into showing us what we're looking for in an input signal, and indeed even simply to display it in a stable fashion. Two of the most basic triggering types are edge triggers and pattern triggers. The latter applies to mixed-signal instruments, allowing users to trigger on a logical combination of analog and digital inputs.

30 October 2013

Oscilloscope Basics: Controlling an Oscilloscope (Part II)

An example of a touch screen-equipped oscilloscope.
Figure 1: An example of a
touch screen-equipped
oscilloscope.
In a recent post, we discussed how to control a modern digital oscilloscope using the front-panel controls. That was a natural place to begin, given that it's the "traditional" means of controlling the instrument and the one that most seasoned users cut their teeth on. But there's more than one way to skin this cat these days. Many of today's oscilloscopes carry touch screens that do everything the front-panel controls can do, plus some things they cannot do.