You need to test, we're here to help.

You need to test, we're here to help.
Showing posts with label source impedance. Show all posts
Showing posts with label source impedance. Show all posts

14 February 2022

Transmission Lines for Oscilloscope Users, Part 4

Figure 1: Characteristic waveform when the source impedance is lower than the cable impedance.
Figure 1: Characteristic waveform when the source
impedance is lower than the cable impedance.
In Part 3, we saw the pattern of reflections that occur when both the source impedance and the oscilloscope input impedance are higher than that of the interconnect, and how those reflections affected the rise time measurement. Now let’s briefly consider what happens when the source impedance is lower than the impedance of the connecting cable. 

For this example, the source voltage is a 3.3 V square wave and the source impedance is 9 Ω. As before, our transmission line is a 50 Ω coaxial cable connecting the source to the oscilloscope. If the oscilloscope input termination is set to 1 MΩ, we see the interesting waveform shown in Figure 1.

29 June 2020

Probe Safety Demystified: CAT Ratings

Measurement Category (CAT) ratings by type of circuit probed.
Figure 1. CAT ratings required to
safely test different electrical sources.
(Sourced from “Measurement Categories,”
Wikipedia, Oct. 28, 2019.)
Any voltage probe will have several published specifications that are meant to indicate under what circumstances that probe is safe to use. They answer questions such as “What is the maximum voltage I can safely run through this probe?” and “Can I safely hold this probe while using it, or does it need to be mounted somewhere far away?” In this post, we’ll explain what the CAT rating is telling you.

CAT ratings are standardized ratings used to categorize the suitability of a voltage measurement device based on the source impedance of what it is used to measure. They are issued by the International Electrotechnical Commission (IEC).

19 December 2018

Using 50-Ohm Coax From DUT to Oscilloscope

A coaxial cable presents high impedance at low frequencies but acts as a transmission line at higher frequencies
Figure 1: A coaxial cable presents high impedance at low
frequencies but acts as a transmission line at higher frequencies
In our recent exploration of 10x passive probes, we've determined that while these types of probes are great general-purpose tools, they're not necessarily going to do the job in specialized measurement circumstances. They're relatively low-bandwidth, low-SNR probes that impose some limitations and, in some scenarios, can deliver potentially misleading or erroneous measurement results if used without clear understanding of their capabilities.