24 October 2018

Secrets of the 10x Passive Probe

The 10x passive probe  becomes a better measurement tool when we understand its limitations
Figure 1: The 10x passive probe
becomes a better measurement
tool when we understand its
limitations
We began this series of posts on oscilloscope probes by putting them in perspective: Probes have a number of different jobs to do, including serving effectively as both a mechanical and electrical interface. Despite having electrical attributes of their own, we want them to grab our signal of interest, but we don't want them to affect that signal in any way.

18 January 2018

How 10X Attenuating Probes Kill Signal-to-Noise Ratio

Figure 1: Signal waveforms captured using a 10X attenuating
probe (top) and a BNC probe (bottom) with tips open
We've begun discussing things that can derail (see what we did there?) your power-rail measurements, such as the deleterious effects of RF interference. In the same context, one should always be mindful of certain characteristics of oscilloscope probes; namely, the 10X attenuating probes that are often lying around on the testbench.