You need to test, we're here to help.

You need to test, we're here to help.
Showing posts with label Smart Trigger. Show all posts
Showing posts with label Smart Trigger. Show all posts

08 November 2021

Finding “Unknown” Waveform Anomalies

Figure 1: Width “exclusion” trigger captures pulse widths outside the range of 980 ns to 1 µs. The first anomaly captured is a width of 2.99 µs.
Figure 1: Width “exclusion” trigger captures
pulse widths outside the range of 980 ns to 1 µs.
The first anomaly captured is a width of 2.99 µs.
Using SmartTriggers® and statistics to find waveform anomalies is easy if you know the characteristics of the anomaly, but how do you find intermittent, anomalous events when you don’t know what you’re looking for? The answer: start with what you do know!  

Look for What’s “Not Normal”

A simple approach is to measure the nominal waveform, then trigger the oscilloscope on waveform elements that differ from nominal. Figure 1 shows a 500 kHz square wave with a roughly 50% duty cycle, so the pulse width is normally about 1 µs (the Width measurement shows a mean of 997 ns). This nominal width does not change significantly with any regularity and gives us a basis on which to begin looking for anomalies.

01 November 2021

Finding Intermittent Events

Figure 1: Statistics for 1261 Width measurements taken over 97 acquisitions on the Measure table. Width statistics can help determine the set up of a Glitch SmartTrigger.
Figure 1: Statistics for 1261 Width measurements
taken over 97 acquisitions on the Measure table.
Width statistics can help determine the set up
of a Glitch SmartTrigger.
Glitches, dropouts, runts, aperiodicity, missed cycles, slow edges—whatever you call them, they are irregular waveform elements that can wreak havoc with you circuit operation. Because they do not occur with regularity, they can be hard to find and correlate with whatever synchronous events may be causing them. How can you use your oscilloscope to easily find intermittent events where they occur? The answer is by judicious application of the oscilloscope’s measurement statistics and SmartTriggers®.

12 April 2021

How to Use Measurement Statistics to Set Up Triggers

Figure 1.  Histogram of the different pulse widths occurring in a pulse-width modulated rectangular pulse train.
Figure 1.  Histogram of the different pulse widths occurring
in a pulse-width modulated rectangular pulse train.
Triggering is an essential element in all modern digital oscilloscopes.  The trigger synchronizes the oscilloscope’s data acquisition with a user-specified event on the signal, be that an edge, threshold crossing or a specific signal characteristic. Teledyne LeCroy Smart Triggers can trigger oscilloscope acquisitions based on properties such as a period, width, low signal amplitude, slew rate or signal loss. These trigger types are ideal for capturing transient events like glitches, but they require knowing at least a range of possible values for the trigger to detect.

Intermittent transient events and glitches are among the most frustrating problems to detect and solve. This is especially true if you have no idea about the nature of the transient. However, you can use the oscilloscope’s measurement tools to help locate these bothersome transients, then use that information to set up your trigger to capture them when they occur. Here’s how.

29 January 2014

Tips and Tricks: Stabilizing a Waveform Trace

One initial challenge to oscilloscope users is achieving a stable display of an input waveform. In almost all cases, the Auto Setup button on Teledyne LeCroy oscilloscopes (and most other modern instruments) will automatically set the oscilloscope's triggering system to get that jumpy trace to settle down.