Figure 1: A typical EMC surge test waveform with 1.2 µs rise time and 50 µs half amplitude time. Zoom trace Z1 shows the details of the rise time. Click any image to expand it. |
Surge pulses are similar to ESD pulses in that they have a very fast rise time, but the fall time is much, much slower. Figure 1 shows a typical surge pulse waveform. Surge testing involves similar measurements to those made for ESD pulse testing—such as rise time, pulse width and max—but surge testing also requires some additional measurements, such as area under the pulse curve and transmitted charge.
Figure 2: To verify the surge generator waveform, the oscilloscope is connected to the generator through an attenuator. |
Following are three, important things to do to make sure you get the best surge measurements from your oscilloscope.
1. Set Attenuation Correctly
As was the case with ESD testing, the oscilloscope is set to a 50 Ω input termination. The input to the oscilloscope from the surge generator has to be less than 5 Vrms. This means that a suitable attenuator must be inserted between the surge generator and the oscilloscope as shown in Figure 2. The attenuation ratio should entered into the Attenuation field on the oscilloscope’s input channel dialog so that the oscilloscope’s readout is calibrated to the properly scaled voltage on the screen display.
2. Set Measurement Thresholds Correctly
Figure 3: The rise time measurement is made between the 10 to 90% signal levels based on the 0 to maximum voltage amplitude, not the default top and base amplitudes. Click any image to expand it. |
The reference levels for the rise time measurement in Figure 3 are shown as dash-dot-dot markers in blue. The high and low percent thresholds appear as dashed blue lines at the 10 and 90% levels. The rise time is about 1.2 µs while the fall time is about 140 µs and maximum voltage is 1.1 kV.
As in the case of the ESD pulse test, we used the optional EMC Pulse Parameter Time to Half Level (EMCt2hv) to measure the time from the leading edge of the pulse until the amplitude falls to half the maximum value, nominally 50 µs.
3. Use Math on Parameters to Correctly Calculate Charge
The surge standards require some measurements in addition to those made for ESD testing. The first is the area under the pulse. Area is a standard oscilloscope measurement and as the name implies uses an integral function to measure the area under the acquired pulse. The result is shown in parameter readout P5 as 76 milli-Webers (Volt-seconds).
The area measurement is then used to calculate the charge transferred by the pulse. If we divide the area by the oscilloscope’s input impedance, we get the charge in Coulombs (Ampere-seconds).
Figure 4: Using Math on Parameters to convert the measured area under the pulse to charge. Both setup dialogs involved are overlaid in the figure. |
Next, the Math on Parameters operator P Ratio is used to divide the area in P5 by the constant in P6. The oscilloscope automatically assigns the proper units, milli-Coulombs (mC), to the calculation, showing a charge of 1.5 mC.
For more EMC surge testing tips, watch our on-demand webinar, "How to Get the Most Out of Your EMC/EMI Lab Oscilloscope Pt. 1".
No comments:
Post a Comment