27 February 2019

About Windowing in Fast Fourier Transforms

Windowing a waveform that's not periodic within the acquisition window reduces spectral leakage
Figure 1: Windowing a waveform that's not periodic within
the acquisition window reduces spectral leakage
The second artifact in FFT calculations is more serious and is also the source of some confusion. A good source for information on this topic of windowing is an application note published by National Instruments.

22 February 2019

About Data Truncation in Fast Fourier Transforms

The first precondition of using the Fourier transform is a repetitive signal
Figure 1: The first precondition of using the Fourier transform
is a repetitive signal
Our last post discussed how time-domain signals acquired by an oscilloscope might be translated into the frequency domain using the discrete Fourier transform. We noted that using the Fourier transform only works if our signal is repetitive (Figure 1), and that it allows us to identify only harmonics of the first harmonic frequency, which is 1/acquisition window. Moreover, the discrete Fourier transform, if used on a large number of data points, is relatively slow to calculate.

30 January 2019

Getting From the Time Domain to the Frequency Domain

A fundamental underlying assumption of a discrete Fourier transform is a repetitive waveform
Figure 1: A fundamental underlying assumption of a discrete
Fourier transform is a repetitive waveform
Fundamentally, oscilloscopes are time-domain instruments: We use them to acquire signals from our circuitry or device under test, and the instrument displays them in the time domain. We see the voltage of the signal in the vertical axis, and we see how that voltage changes over time in the horizontal axis.

19 December 2018

Using 50-Ohm Coax From DUT to Oscilloscope

A coaxial cable presents high impedance at low frequencies but acts as a transmission line at higher frequencies
Figure 1: A coaxial cable presents high impedance at low
frequencies but acts as a transmission line at higher frequencies
In our recent exploration of 10x passive probes, we've determined that while these types of probes are great general-purpose tools, they're not necessarily going to do the job in specialized measurement circumstances. They're relatively low-bandwidth, low-SNR probes that impose some limitations and, in some scenarios, can deliver potentially misleading or erroneous measurement results if used without clear understanding of their capabilities.

12 December 2018

Squeezing More Bandwidth From a 10x Passive Probe

Shown is a comparison of inherent oscilloscope noise and noise at the shorted tip of a 10x passive probe
Figure 1: Shown is a comparison of inherent oscilloscope
noise and noise at the shorted tip of a 10x passive probe
Now that we have a better understanding of what's happening under the hood of a 10x passive oscilloscope probe, we can sum up its key characteristics. The first thing to know about such probes is that they offer relatively low bandwidth (<100 MHz). This is largely a result of the probe's tip inductance.

28 November 2018

10x Passive Probes and Cable Reflections

Figure 1: With unequal impedances at either end of the coax, are cable reflections a concern in 10x passive probes?
Figure 1: With unequal impedances at either end of the coax,
are cable reflections a concern in 10x passive probes?
We've been discussing the ubiquitous 10x passive probe here on Test Happens, beginning with an overview of the probe-oscilloscope system. We turned to the 10x passive probe itself and the issues posed by its constitutive circuitry. Then we covered what about that circuitry makes it usable at all, namely, its built-in equalization circuit.

08 November 2018

How Tip Inductance Impacts a Probing System's Bandwidth

Shown are FFT plots of a 10-MHz, fast-edge square wave reaching the oscilloscope via direct coax connection  (orange-yellow plot) and 10x passive probe fitted with a coax tip adapter (straw-colored plot)
Figure 1: Shown are FFT plots of a 10-MHz, fast-edge square
wave reaching the oscilloscope via direct coax connection
(orange-yellow plot) and 10x passive probe fitted with a coax
tip adapter (straw-colored plot)
If you're using 10x passive probes with your oscilloscope, it's important to understand the bandwidth of your probing system and how it's affected by various methods of probing the signal of interest. There's a relatively easy way to determine this parameter by probing a fast-edge, 10-MHz signal from a square-wave generator. Doing so can also instruct us in the effects of tip inductance on the probe's bandwidth.